Isolated Stimulation

World Precision Instruments Inc

The term stimulation refers to the delivery of energy of some kind to a biological tissue in order to elicit an observable response.

Isolated Stimulation and Stimulus isolators

The term stimulation refers to the delivery of energy of some kind to a biological tissue in order to elicit an observable response.

Although the energy used in stimulation may be chemical, thermal, mechanical or electrical, this discussion will focus on electrical stimulation. Electrical stimulation of biological tissues involves the delivery of current and voltage to the stimulation site. The two quantities are related by Ohm's law:

V=IR

Where V is the applied voltage, I is the current and R is the electrical resistance of the tissue and or the stimulating electrodes. This simple equation shows that if voltage is constant, current flow will diminish if the tissue/electrode resistance goes up and will increase if the resistance decreases.

More commonly, the resistance of tissue differs from sample to sample, and the resistance of the electrodes changes with applied current over time in a process called polarization.

Types of Stimulus Devices
In the delivery of electrical energy to biological tissues, stimulus devices can hold either current or voltage constant during the stimulating process. Devices that hold voltage constant at a value set by the user and allow current to be determined by Ohm's law are known as constant voltage stimulators.

Devices that hold current constant at a value set by the user during the stimulation process and allow voltage to be determined by Ohm's law are called constant current stimulators.

Constant current stimulators are preferred for two reasons:

  • First, current is the quantity that stimulates most excitable tissues.
  • Second, stimulating electrodes tend to increase their resistance as stimulation progresses, as do some tissues. A constant current stimulator will "sense" resistance change and provide whatever voltage is needed to maintain delivery of current at the set rate.

There is obviously a limit to how much voltage a constant current stimulator can provide. If the resistance of the preparation were to become infinite, as might happen if one of the stimulating electrodes was removed from the tissue, the stimulator could not mount an infinite voltage to compensate.

The maximum amount of voltage that a constant current stimulator can provide is called the compliance voltage. Once this compliance voltage has been reached, further increases in tissue resistance will cause a drop in delivered current.

WPI isolators in the 300 and DLS series' offer a compliance voltage of 100V with very low noise.
Stimulus isolators, as the name implies, also isolate a given stimulus from ground. In an instrument design context most people think of isolation from ground as it relates to electrical safety. From a biological recording standpoint there are other issues. Consider the circuit in Figure 1.

Stimulus Devices

The stimulator in this case is a battery with a switch; current leaves the positive terminal of the battery, travels down the stimulating lead, passes through the tissue, and 100% of the current delivered returns to the negative terminal of the battery. This figure also shows a voltmeter in the preparation.

Voltages generated by the tissue as a consequence of the stimulus are recorded with respect to the voltmeter's ground electrode.

Line Powered Stimulator

In Figure 2 the battery is replaced with a line powered stimulator. Even though the stimulus source and the voltmeter have separate ground electrodes, they represent the same electrical point. For this reason a significant portion of the stimulus current returns to ground by way of the voltmeter's ground lead.

If the currents are significant, or the voltages that you are trying to measure are very small, the I x R drop across the resistance of the voltmeter's ground electrode will add to the recorded voltage from the tissue and will be seen as a DC artifact.

Capacitative coupling between the voltmeter circuit and the isolated circuit can induce current to flow in the voltmeter ground. The induced current will transiently flow across the resistance of the voltmeter ground and its I x R drop will be seen as a transient spike in front of and behind any pulse of current delivered by the current source. This is termed an AC artifact.

The primary reason why researchers use an isolated current source is to minimize artifact. But what if you are not recording? There is no voltmeter ground to produce an observable artifact. If you stop and consider the examples above, the artifact was minimized or eliminated by controlling the path of the stimulus current.

Knowing the current path can be critical in physiological stimulation. Consider the current path in Figure 3. The animal represented by the badly drawn cat is secured in a stereotaxic frame.

Isolated Stimulation / Stimulus Isolators

The frame is grounded. The animal contacts the frame at multiple points. Figure 3 shows the battery and switch model of a stimulator. Here as before all of the current that leaves the positive side of the battery must return to the negative side. 100% of the stimulus current must pass between the stimulus electrodes. In this case the path as well as the exact amount of current delivered is known.

Isolated Stimulation / Stimulus Isolators

Figure 4 shows the same experimental setup except the battery has been replaced by a line powered voltage source. Stimulus current now returns to ground via many routes. The amount of current that flows to ground is determined by the resistance between the source and each of the ground points and is calculated as a resistive network using Kirtchoff's laws. Many unintended areas of the animal may be stimulated.

In the real world we cannot use a battery and a switch, particularly if the current durations are on the order of milliseconds. Electronic devices such as pulse generators and computers are used to generate timing, and Isolators driven by these devices are used to deliver the stimulus.

By connecting an isolator (even a battery powered one) to a mains powered pulse generator, you connect the isolated ground of the isolator to the mains ground of the pulse generator. Unless the electrical connection between the two devices is accomplished without using a mechanical connection between the two devices you break the isolation.

This is what makes an isolator an isolator. The non-mechanical contact between machines that constitutes the isolation barrier can be accomplished in one of two ways.

Originally, isolators were transformer isolated: pulse waves were applied to the primary winding of a transformer, while the actual stimulus was derived from the secondary winding. The transference was accomplished by induction. This approach suffered from two shortcomings.

The devices could not pass DC so no constant voltage isolated stimulations could be made. The transformer approach also has intrinsically higher capacitance. This means that while the resistance between the primary and secondary coils is very high, the high capacitance will produce an AC artifact that is unacceptably large compared to other isolation techniques.

Optical isolation is the second popular scheme. The field has all but standardized on this. In simple terms, the input pulse wave powers a light that shines across the barrier onto a photocell that produces the stimulus wave. There have been countless variations on this theme and it is used for isolating recording amplifiers as well as stimulators.

WPI's Isolators, all of which are optical:

DLS Series

WPI’s Isolators, all of which are optical:WPI's digital stimulator, DS8000 has advanced features and specifications that are not found in any other digital stimulator.

The DLS100 are a revolutionary new breed of digital linear isolators with high compliance and high isolation.

The DLS100 is preferentially optimized for applications in which the DS8000 digital stimulator is used and connects to the DS8000 via a flexible cable through which it receives power and stimulus signals in a digital format.

Up to eight DLS100 isolators can be connected independently to one DS8000.

Very high isolation is achieved through the use of optical coupling of the digital signal and a galvanically isolated DC power supply within the DLS100.

Unlike some other multi-channel isolators, this digital isolator can be located at the site of the experiment, allowing the use of short connecting leads and thereby preserving high isolation and fast signal rise and fall times.

The DLS100 operates in two modes: current source or voltage source. In the current/voltage source mode, the output current/voltage is proportional to the amplitude and polarity of the signal generated by the DS8000. Over-range can occur when the resistance of the load (the experiment) is too high for the current or voltage that is demanded from the DLS100.

DLS100 Specifications

Current Source Mode

 

Full-scale* Current

10 mA, 100 µA, 10 µA, bipolar

Compliance Voltage

± 100 volts

Output Impedance

Greater than 100 Megohms

Zero-signal Leakage

Less than 0.01% of full-scale range setting

Linearity

Better than 0.05% of full-scale range setting

Bandwidth

Range and load dependant: 20 kHz with 10K load and 100 µA or above range.**

Voltage Source Mode

 

Full-scale* Voltage

± 100 volts

Max. Current

1o mA

Output Impedance

Less than 1 ohm

Zero-signal leakage

Less than 1 mV

Linearity

Better than 0.05% of full-scale range setting

Bandwidth

50 kHz

Isolation

.

Resistance

Greater than 1000 Megohms

Capacitance

Less than 10 pF, from output terminals to DS8000 and earth ground

Power Requirements

+ 12 volts and +5 volts, supplied by DS8000

Output Terminals

Mini-banana jacks


Activated by conventional logic-level commands, Model A360LA can be gated by any pulse generator, stimulator, or computer output. It has dual tone audible alarms and a test switch is also provided to check battery charge. Model A360LA also has stimulus currents that can be set using a three-digit control knob and a three-position range switch. The voltage sufficient to push the desired current through the load is automatically developed, subject only to compliance limits.

Isolated Stimulation / Stimulus Isolators

A360LA SPECIFICATIONS

Output Waveform

Dc Or Current Pulse

Output Current Ranges

0.1, 1.0, And 10 Ma

Current Amplitude Error

0.5% Of Full Scale, Max.

Current Resolution

0.1% Of Full Scale, Typical

Output Load Voltage Excursion (Compliance)

100 V Minimum, 130 V Nominal

External Command Threshold

2.2 V At 2.6 Ma, Min. 8.5 V, Max. (Ttl)

Output Polarity

Reversible, Manual Switch

Current Rise Time & Delay

2 µS, Typical (1 Kω Load) 5 µS @ 10 Kω

Current Fall Time & Delay

5 µS, Typical (1 Kω Load) 8 µS @ 10 Kω

Output To Ground Resistance

1012Ω

Optocoupler

2500 V, Rated Min. Breakdown Voltage

Noise

<0.1 Microa (10 Ma), < 50 Na (1 Ma, 100 Microa)

Current Leakage

+/- 10 Na +/- 0.01% Of Full Scale

Power

Two 6-V Lead/Acid Batteries, 2.4 Ah

Recharge Time

4 Hours

Recharger Included

100-240 Vac, 50-60 Hz, Universal


Model A365 includes the same features and specifications as A360 but with the added capability for bipolar pulsing for zero net charge. The bipolar current is toggled by the command waveform, setting alternating pulses as positive or negative.

Isostim™Stimulator/Isolator can now come with NEW NiMH Batteries in a rechargeable version. It combines the ease of use and accuracy of WPI's 300 Series stimulators with the power output of a stimulus isolator. External/DC mode converts Isostim™ to a passive stimulus isolator.

The A320D model is powered by readily obtainable 9-volt alkaline batteries (included). The rechargeable A320R is supplied with a nickel cadmium battery stack that provides 10-12 hours of operation before recharge is required. The A362 battery charger must be used with the A320R.

WPI also offers high current stimulus isolator, which combines optical isolation with a ±100 mA current generator. A365 model delivers positive, negative, or bipolar currents. The input connector is a standard BNC, allowing signals from any source - such as computer D/A or I/O lines - to be used.

A385 rechargeable version is not appropriate for transcutaneous stimulation, and the A382 system charger must recharge the batteries. Indicator lights and audible alarms keep the user constantly aware of battery charge status.

WPI's linear stimulus isolator model A395, also available in rechargeable version will replicate a programmed waveform of any shape or polarity. Battery operated, and photoelectrically-isolated from the input voltage drive, the instrument regenerates as output currents the original waveforms provided by your D/A converter or signal generator.

RSS