Conductivity : Industrial Online Conductivity Monitor / Controller

Russell Mainstream Supply Ltd

The Alpha CON2000 is our top of the range industrial Conductivity process control instrument that is proving to be very successful in applications involving agriculture, chemical product, boiler feed water, water treatment, effluent treatment, waste water, pharmaceuticals, pulp and paper industries and bleach manufacturing.

The Alpha Series 2000 1/2 DIN controllers, can be wall-mounted or panel-mounted, are equipped with advanced menu-driven incorporated with a myriad of user-customizable features. Features include built-in PI/proportional control, scaleable 4 - 20 mA current output for fine control, 2 independent set-points, alarm with timer control, programmable wash cycle,

Conductivity MonitorApplications for the Alpha CON2000 include:

General:

Use for virtually any batch or online applications where rapid, accurate control. Great for OEM/system integrator.

Industrial:
Use in applications involving agriculture, chemical processing, boiler feed waters, manufacturing, pharmaceuticals, pulp and paper industries, and bleach manufacturing.

Water Purification/Treatment:
Use to treat batches of incoming process water, ultrapure water, boiler and feed water control.

Regulatory:
Hook to recorder to document data for regulatory compliance.

Main Features:

Seven ranges in a single controller allows the controller to be used in a wide variety of applications with the maximum range at 1000 mS/cm

User-Customization through Advanced Setup Menu offers flexibility in matching the controller's functions to suit individual's specific requirement

Choice of 4-cell or 2-cell electrode system

Programmable for On/Off, Proportional, or PI (Proportional Integral Control) - ideal for most demanding precision process control applications

Backlit Display offers reading in low-light conditions

Universal Power Supply Input from 85 to 250 VAC/DC allows use at multiple sites and withstands voltage fluctuations

Two Galvanically Isolated Scaleable 0-20/4-20 mA Outputs for Conductivity and temperature

Dual Display shows Conductivity with temperature simultaneously - features clear multiple icons, set points, and status messages

Two Galvanically Isolated Scaleable 0-20/4-20 mA Outputs

Wash Contact Relay controls electrodes cleaning systems at desired duration and frequency

Adjustable Hysteresis (Dead Band) prevents rapid contact switching near set point

Non-Volatile Memory retains all stored parameters and calibration data even if power fails

Large Dual Display shows measurement with temperature simultaneously - features clear multiple icons, set points, and status messages

Choice of Temperature Sensor Pt100/Pt1000 with 2-wire or 3-wire temperature input selection

Rugged NEMA 4X housing simplifies installation outdoors and indoors. Pipe, Wall or Panel mount

Easy Installation and Wiring with detachable plug-in connectors

Panel-mounted version is also available

Specifications:

Product CodeEC-CONCTP2000W / EC-C0NCTP2000P
Conductivity Range0.000 - 2.000, 20.00, 200.0, 2000 µS/cm;
20.00, 200.0, 1000 mS/cm
Resolution & Accuracy0.001, 0.01, 0.1, 1 µS/cm; 0.01, 0.1, 1 mS/cm
& ± 1% Full Scale (± 2% Full Scale > 500 mS/cm)
Relative Accuracy±1% of full scale reading (±2% > 500 mS/cm)
Temperature Range-10.0 to +125.0°C (14.0 to 257.0°F)
Resolution0.1°C/°F
Relative Accuracy±0.5°C (±0.9°F)
SensorPt100/ Pt1000 (jumper selectable); 2 or 3 wire
Cell Constants0.01, 0.1, 1.0, 10.0
Temperature CompensationAuto/ manual (normalised at 25°C)
Type of Controller SettingP/PI, Limit, Off (Monitoring);
Proportional (Pulse Length or Pulse Frequency)
No. of Relays
Maximum Voltage
Maximum Current
High/Low Hysteresis Band
3 SPDT; 1 Alarm
250 VAC
3A;
0 to 10% Full Scale
Pickup/ Dropout Delay0 to 2000 seconds
Wash Cycle/Duration0.1 to 199.9 hrs & 1 to 1999 sec
Power80 to 250 VAC/DC 50/60 Hz Approx. 10VA
Dimensions1/2 DIN (144 x 144 mm)
RSS